OCTAX

External

| |
Some trouble shooting information for typically used microscopes when used
with the OCTAX Laser Shot™ in ART
Olympus
IX50
- No laser interaction: For Laser Shot™ systems with turret: Shutter
must be open (see symbol) and the turret in the indicated position
- Horizontal misalignment of the monitor image: beam splitter switch
between "only eyepiece" and "side port and eyepiece" must be locked in the
correct position (the switch must not be in a position between the two
settings)
IX70
- Horizontal misalignment of the monitor image: beam splitter wheel
(knob) between "only eyepiece" and "side port and eyepiece" must be in locked
the correct position (the switch must not be in a position between the two
settings), i.e. one of the two possible end positions of the wheel, on the
right side of the microscope
- Misalignment of the monitor image: while using the laser, the 1.5x
magnifier below the objective revolver on the right side must be completely in
in position (1x) and not in position (1.5x) or an intermediate position
IX51 and IX71
- Horizontal misalignment of the monitor image: beam splitter switch
between "only eyepiece" and "side port and eyepiece" must be locked in the
correct position (the switch must not be in a position between the two
settings)
- (IX 71 only) Misalignment of the monitor image: while using the
laser, the 1.5x magnifier below the objective revolver on the right side must
be completely in in position (1x) and not in position (1.5x) or an
intermediate position
Nikon
Diaphot 200
- Horizontal misalignment of the monitor image: beam splitter wheel
(knob) between "only eyepiece" and "side port and eyepiece" must be in locked
the correct position (the switch must not be in a position between the two
settings)
Diaphot 200
- Horizontal misalignment of the monitor image: beam splitter wheel
(knob) between "only eyepiece" and "side port and eyepiece" must be in locked
the correct position (the switch must not be in a position between the two
settings)
Te200
- Horizontal misalignment of the monitor image: beam splitter wheel
(knob) between "only eyepiece" and "side port and eyepiece" must be in locked
the correct position (the switch must not be in a position between the two
settings)
Te300
- Horizontal misalignment of the monitor image: beam splitter wheel
(knob) between "only eyepiece" and "side port and eyepiece" must be in locked
the correct position (the switch must not be in a position between the two
settings)
Te2000s
- No laser interaction: For Laser Shot™ systems with turret: Shutter
must be open (see symbol) and the turret in the indicated position
- Horizontal misalignment of the monitor image: beam splitter wheel
(knob) between "only eyepiece" and "side port and eyepiece" must be in locked
the correct position (the switch must not be in a position between the two
settings)
Te2000u
- No laser interaction: For Laser Shot™ systems with turret: Shutter
must be open (see symbol) and the turret in the indicated position
- Horizontal misalignment of the monitor image: beam splitter wheel
(knob) between "only eyepiece" and "side port and eyepiece" must be in locked
the correct position (the switch must not be in a position between the two
settings)
- Vertical misalignment of the monitor image: while using the laser,
the 1.5x magnifier below the objective revolver on the right side must be
completely locked in in position (1x) and not in position (1.5x)
or an intermediate position
Can't find your microscope?
Contact your distributor or OCTAX
support.
|